Episode 918: Automated Blood Pressure Cuffs

Episode 918: Automated Blood Pressure Cuffs

Contributor: Aaron Lessen, MD

Educational Pearls:

How does an automated blood pressure cuff work?

  • Automated blood pressure cuffs work differently than taking a manual blood pressure.

  • While taking a manual blood pressure, one typically listens for Korotkoff sounds (turbulent flow) while slowly deflating the cuff.

  • An automatic blood pressure cuff only senses the pressure in the cuff itself and specifically pays attention to oscillations in the pressure caused by when the pressure of the cuff is between the systolic (heart squeezing) and diastolic (heart relaxed) pressures.

  • These oscillations are at a maximum when the pressure in the cuff matches the mean arterial pressure (MAP) and therefore the machines are most accurate at reporting the MAP.

  • The machines then use the MAP and other information about the oscillations to estimate the systolic and diastolic pressures, which are less accurate.

What should you do if you need more accurate systolic and diastolic blood pressures?

  • Take a manual blood pressure.

  • Get an arterial-line (a-line), which provides continuous data for the blood pressure at the end of a catheter.

What happens if the cuff is too big or too small for the patient?

  • If the cuff is too small it will overestimate the pressure.

  • If the cuff is too large it will underestimate the pressure.

What should you do if the cuff cycles a bunch of times before reporting a blood pressure?

  • It probably isn't very accurate so consider another method.

Bonus fact!

  • The MAP is not directly in the middle of the systolic and diastolic pressures but is weighted towards the diastolic pressure. The MAP can be calculated by adding two-thirds of the diastolic pressure to one third of the systolic pressure. For example if the BP is 120/90 the MAP is 100 mmHg.

References

  1. Benmira, A., Perez-Martin, A., Schuster, I., Aichoun, I., Coudray, S., Bereksi-Reguig, F., & Dauzat, M. (2016). From Korotkoff and Marey to automatic non-invasive oscillometric blood pressure measurement: does easiness come with reliability?. Expert review of medical devices, 13(2), 179–189. https://doi.org/10.1586/17434440.2016.1128821

  2. Liu, J., Li, Y., Li, J., Zheng, D., & Liu, C. (2022). Sources of automatic office blood pressure measurement error: a systematic review. Physiological measurement, 43(9), 10.1088/1361-6579/ac890e. https://doi.org/10.1088/1361-6579/ac890e

  3. Vilaplana J. M. (2006). Blood pressure measurement. Journal of renal care, 32(4), 210–213. https://doi.org/10.1111/j.1755-6686.2006.tb00025.x

Summarized by Jeffrey Olson, MS3 | Edited by Meg Joyce, MS1 & Jorge Chalit, OMS3

Donate: https://emergencymedicalminute.org/donate/

Avsnitt(1146)

UNTAMED PART 2: Surviving a Medical Emergency in the Outdoors

UNTAMED PART 2: Surviving a Medical Emergency in the Outdoors

11 Mars 201917min

UNTAMED PART 1: The Physiology of Experiencing Nature

UNTAMED PART 1: The Physiology of Experiencing Nature

11 Mars 201917min

Podcast #445: Hunting for the endotracheal tube

Podcast #445: Hunting for the endotracheal tube

Author: Michael Hunt, MD Educational Pearls: Bedside transtracheal ultrasound to confirm proper endotracheal intubation is simple and effective Review of 17 studies showed transtracheal ultrasound wa...

8 Mars 20192min

Podcast #444: Dyspnea in the intubated patient

Podcast #444: Dyspnea in the intubated patient

Author: Aaron Lessen, MD Educational Pearls: Dyspnea can occur in up to 50% of intubated patients Dyspnea is associated with prolonged mechanical intubation Often goes undiagnosed as these patients h...

6 Mars 20192min

Podcast #443: Measles - a timely revisit

Podcast #443: Measles - a timely revisit

Author: Sue Chilton, MD Educational Pearls: Measles is highly contagious Typically patients just look sick Remember 4 & 4, 14, and 4-C's: Contagious 4 days before and after onset of rash 14 days fo...

4 Mars 20193min

Podcast #442: When a severe man cold isn't

Podcast #442: When a severe man cold isn't

Author: Jared Scott, MD Educational Pearls: CXR is 40-90% sensitive for detecting pneumonia when compared to CT Patients with a high degree of suspicion for pneumonia may still warrant treatment de...

2 Mars 20194min

Podcast #441: Is Heparin Worthless for ACS?

Podcast #441: Is Heparin Worthless for ACS?

Author: Don Stader, MD Educational Pearls: Recent study demonstrated patients with non-STEMI treated with heparin had no difference in survival, but had an increased risk of major bleeding Heparin ...

27 Feb 20193min

Podcast # 440 : Carbon Monoxide Poisoning

Podcast # 440 : Carbon Monoxide Poisoning

Author: Katie Sprinkle, MD Educational Pearl: Carbon monoxide (CO) is the leading cause of death from poison world-wide CO forms from the combustion of any carbon based product Patients often prese...

25 Feb 20194min

Populärt inom Vetenskap

dumma-manniskor
svd-nyhetsartiklar
p3-dystopia
allt-du-velat-veta
kapitalet-en-podd-om-ekonomi
det-morka-psyket
rss-ufo-bortom-rimligt-tvivel
rss-vetenskapsradion
pojkmottagningen
sexet
rss-vetenskapsradion-2
dumforklarat
rss-broccolipodden-en-podcast-som-inte-handlar-om-broccoli
medicinvetarna
halsorevolutionen
bildningspodden
rss-experimentet
rss-spraket
hacka-livet
vetenskapsradion